
140
Chapter 4 Features and Functions
Alarm Limits
• You can assign an alarm to any configured channel and multiple
channels can be assigned to the same alarm number. However, you
cannot assign alarms on a specific channel to more than one alarm
number.
• When an alarm occurs, the instrument stores relevant information
about the alarm in the queue. This includes the reading that caused
the alarm, the time of day and date of the alarm, and the channel
number on which the alarm occurred. The information stored in the
alarm queue is always in absolute time format and is not affected by
the
FORMat:READing:TIME:TYPE
command setting.
• You must configure the channel (function, transducer type, etc.)
before setting any alarm limits. If you change the measurement
configuration, alarms are turned off and the limit values are cleared.
Alarms are also turned off when you change the temperature probe
type, temperature units, or disable the internal DMM.
• If you plan to use alarms on a channel which will also use scaling,
be
sure to configure the scaling values first
. If you attempt to assign the
alarm limits first, the instrument will turn off alarms and clear the
limit values when you enable scaling on that channel. If you specify a
custom measurement label with scaling, it is automatically used
when alarms are logged on that channel.
• If you remove a channel from the scan list (by selecting
CHANNEL OFF
from the front panel or by redefining the scan list
from the remote interface), alarms are no longer evaluated on that
channel (during a scan) but the limit values
are not
cleared. If you
decide to add that channel back to the scan list (without changing the
function), the original limit values are restored and alarms are
turned back on. This makes it easy to temporarily remove a channel
from the scan list without entering the alarm values again.
• Each time you start a new scan, the instrument clears all readings
(including alarm data) stored in reading memory from the previous
scan. Therefore, the contents of reading memory are always from the
most recent scan.
34970A Refresh UG.book Page 140 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...