91
Chapter 5: Troubleshooting
Troubleshooting the Agilent 1680A,AD-series
b
As the instrument is booting, probe for digital signals on the hard disk
drive connector according to the following table.
Disk Drive Voltages
6
Troubleshoot a flexible disk drive.
a
Apply power to the instrument.
b
After the instrument finished booting, launch the Agilent Logic
Analyzer application.
c
Insert a formatted flexible disk in the instrument flexible disk drive.
d
Attempt to do a File Save of the Agilent Logic Analyzer default
configuration to the flexible disk drive.
Pin No.
Signal
Voltage
Pin No.
Signal
Voltage
Pin No.
Signal
Voltage
1
RESET
27
IORDY
34
PDIAG
3 - 18
DATA
28
CSEL
35
DA00
20
KEY
29
DMACK
36
DA02
21
DMARQ
31
INTRQ
37
CS0
23
DIOW
32
IOCS16
38
CS1
25
DIOR
33
DA01
39
DASP
Pins 2, 19, 22, 24, 26, 30, 40, are GROUND
Pins 41 and 42 are +5 Vdc