41
Chapter 3: Testing Performance
To test the single-clock, single-edge, state acquisition
c
If the pulse width is outside the limits, adjust the pulse generator
channel 2 width until the pulse width is within limits.
Check the setup/hold combination
The following setup/hold combinations will be tested:
Setup/Hold Combinations
1
Disable the pulse generator channel 1 COMP (with the LED off).
2
Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup time of the setup/hold combination selected,
+0.0 ps or
−
100 ps as measured on the oscilloscope:
a
On the Oscilloscope, select [Define meas] Define
∆
Time - Stop edge:
rising, Edge number 2.
b
In the oscilloscope timebase menu, select Position. Using the
oscilloscope knob, position the data waveform so the falling edge is near
the center of the display.
Test
Combination
Setup/Hold
Times
Setup/Hold
Window
Sample Position
(in middle of Window)
1
4.50/-2.0 ns
2.5 ns
-3.25 ns
2
-2.0/4.50 ns
2.5 ns
+3.25 ns
Summary of Contents for 1680 series