85
Chapter 5: Troubleshooting
Troubleshooting the Agilent 1680A,AD-series
Uninstall, then reinstall
the Agilent Logic
Analyzer application.
Does the
application
software launch?
Do error
messages appear?
Perform "To recover
the operating system"
in Chapter 5.
No
Yes
No
Yes
Remove power from
instrument, remove
cover, check internal
IEEE 1394 cabling.
Are all cables
connected and
seated?
Reconnect or reseat
all internal cabling.
Possible problem with
acquisition board.
Replace acquisition bd.
Reseat the PCI IEEE
board in ISA slot 1*.
Apply power to the
instrument.
* All PCI boards must be installed in
the correct slots. See "To remove
the ISA board" in Chapter 6.
Is the application
Online?
Observe the status LEDs
on the acquisition board.
Refer to Acquisition board
status LEDs.
Are any green
LEDs blinking?
Possible problem with PCI
IEEE 1394 board. Replace
the PCI IEEE 1394 board.
Yes
No
Yes
No
Yes
No
6A
1
6B
1
Summary of Contents for 1680 series