42
Chapter 3: Testing Performance
To test the single-clock, single-edge, state acquisition
c
On the oscilloscope, select [Shift]
∆
Time, then select [Enter] to display
the setup time (
∆
Time(1)-(2)).
d
Adjust the pulse generator channel 1 Delay until the pulses are aligned
according to the setup time of the setup/hold combination selected,
+0.0 ps or –100 ps.
3
Select the logic analyzer sample positions:
a
Click the
Sampling Setup icon. The Analyzer Setup dialog opens with
the Sampling tab displayed.
b
Click Sample Positions....
c
In the Sample Positions dialog, drag the blue bar for My Bus 1 to the
sample position of the first setup/hold combination to be tested (or
enter the value in the signal fields).