88
Chapter 5: Troubleshooting
Troubleshooting the Agilent 1680A,AD-series
3
Remove the power supply from the instrument. Refer to “To remove the
power supply” in Chapter 6.
4
After removing the power supply, connect a power cord to the power
supply and plug the power cord into line power.
5
Using DVM, measure the power supply voltages.
Power Supply Voltages
6
Note problems with the power supply, then unplug the power supply from
line power. Return to the flow chart.
CN1
CN2
Pin
Voltage
Pin
Voltage
1-5
+3.3 V
1-4
-5.2 V
6-7
COM
5
+12 V
8-10
+5 V
6-8
-12 V
11-12
COM
9-12
COM
13-14
+3.3 V
13
+12 V
15-19
COM
14-16
COM
20-21
+5 V
22-24
COM