76
BOD Setup:
BOD Configuration
XL40 and 60 meters
This screen allows you configure the BOD seed minimum Delta DO, seed minimum DO
endpoint, sample minimum Delta DO and sample minimum DO endpoint. You may set each
from 0.0 to 60.0.
To set BOD Configuration
1. Touch the
Touch here to edit
button next to the BOD Configuration to access the BOD
Configuration screen.
2. Touch the
Seed Min Delta D.O.
box and use the numeric keypad on the screen to enter
desired value. The
BS
key allows you to remove characters that are incorrectly entered.
3. Touch
Enter
on the keypad to accept the value and return to the BOD Configuration
screen.
4. Repeat steps
2
and
3
for the remaining boxes, Seed Min D.O. Endpoint, Sample Min Delta
D.O. and Sample Min D.O Endpoint.
5. Touch
OK
button to accept the changes of the entire BOD configuration and return to the
BOD Setup screen.
The
Seed Minimum Delta DO
refers to the difference between the initial and final DO
measurements for the seed used. If the delta is less than the minimum value entered, that BOD
value will not be used to calculate seed values. The seed must exhibit a final DO value greater
than the
Seed Minimum DO Endpoint
value entered. If not, it will not be used in computing seed
values.
The
Sample Minimum Delta DO
refers to the difference between the initial and final value DO
measurements of the sample. If the difference is less than the minimum value entered, that BOD
value will not be used to calculate average BOD measurements. The sample must exhibit a final
DO value greater than the
Sample Minimum DO Endpoint
value entered. If not, it will not be
used in computing average BOD values.
Summary of Contents for XL 20
Page 1: ......
Page 138: ...127 Ion Operation Known Addition Method XL25 50 and 60 meters...
Page 140: ...129 Ion Operation Known Subtraction Method XL25 50 and 60 meters...
Page 142: ...131 Ion Operation Analate Addition Method XL25 50 and 60 meters...
Page 144: ...133 Ion Operation Analate Subtraction Method XL25 50 and 60 meters...