56
Ion Setup: Set
Default Temperature
XL25, 50 and 60 meters
Set Default Temperature
It is a well known fact that ion concentration is a temperature dependent measurement. The
factory default setting is 25°C. If you are measuring a solution that is not 25°C and if you are not
using a temperature probe, then you should enter the temperature value of that solution in order
to get the correct concentration value. The current default temperature setting will be displayed
on the screen.
The default temperature can be set from -5°C to 105°C.
To set Default Temperature
1. Select temperature units by touching the appropriate unit button:
C
(Celsius),
F
(Fahrenheit) or
K
(Kelvin).
2. Touch the
Default Temperature
box and use the numeric keypad to enter the desired
default temperature (-5°C to 105°C).
3. Press
Enter
in the keypad to return to Ion Setup screen.
Best results are obtained when the standards and the samples are at the same temperature.