124
Ion Operation:
Direct Reading Methods
XL25, 50 and 60 meters
Measuring Ion Concentration in the Direct Reading with Blank Offset Mode
With Auto Read on
1. Add ionic strength adjuster to the samples.
2. Immerse the ion selective electrode in the stirring sample
3. Touch
Measure.
When the meter recognizes the reading as stable,
STABLE*
will appear
on the screen. The value that appears on the screen is the value of the sample minus the
value of the blank reading. The measure screen will fix on this measurement and will not
deviate until
Measure
is touched again.
4. Repeat steps 1-3 with additional samples. The meter will update the reading each time
Measure
is touched.
* Please note that
STABLE
will only appear if you have completed a successful two point calibration.
With Auto Read off
1. Add ionic strength adjuster to the sample.
2. Immerse the ion selective electrode in the stirring sample. The meter will continually
monitor the ion concentration value. When the meter recognizes the reading as stable,
STABLE
is displayed on the measure screen. With the
Auto Read
function deactivated,
the display on the measure screen will indicate any deviation in the concentration reading.