68
DO Setup:
Select Temperature Units / Select Barometer
XL40 and 60 meters
Select Temperature Units
This setup function allows you to choose the temperature units show in the measurement screen
(celsius, fahrenheit or Kelvin).
To select Temperature Units
1. Select temperature units by touching the appropriate unit button:
C
(Celsius),
F
(Fahrenheit) or
K
(Kelvin).
Set Barometer
The XL40 and XL60 meters use an internal barometric pressure sensor in both the auto and
manual calibration mode. Prior to calibration check the actual barometric pressure. If this value is
different than the current barometric pressure displayed on the XL40 and XL60, you must
calibrate the internal barometer. See below to set the barometer.
DO NOT
use the barometric
pressure issued by the weather service since this value is adjusted to sea level.
The barometer can be set from 450 to 825 mmHg or 599 to 1099 mbar.
To set Barometer
1. Touch
mmHg
or
mbar
to choose desired barometric units.
2. The current internal barometric pressure value is displayed on the
Barometer
box.
3. Touch the
Barometer
box and use the numeric keypad on the screen to enter the actual
barometric pressure. The
BS
key allows you to remove characters that are incorrectly
entered.
4. Touch
Enter
on the keypad to accept the barometric pressure value and return to the
Setup screen.
Summary of Contents for XL 20
Page 1: ......
Page 138: ...127 Ion Operation Known Addition Method XL25 50 and 60 meters...
Page 140: ...129 Ion Operation Known Subtraction Method XL25 50 and 60 meters...
Page 142: ...131 Ion Operation Analate Addition Method XL25 50 and 60 meters...
Page 144: ...133 Ion Operation Analate Subtraction Method XL25 50 and 60 meters...