82
The label on your test
strip container shows the
acceptable ranges for both
the Level 1 and the Level 2
control solutions. The test
result you get should be inside
this range. Ensure you select
and compare the test result to
the correct level of control. For
the level selected, the meter
displays whether the result
is within or outside of the
acceptable range.
When the control test result
is inside the range on the test
strip container, the test strips
and the meter are working
properly.
If the control test result is
within the acceptable range,
“Within range” appears on the
detailed Control Result screen.
Example
Range (mmol/L)
Level 1 1.4-3.1
Level 2 14.2-19.1
Example Control Results
53895_00048010161_E.indb 82
9/4/15 9:34 AM