Display Shows
Possible Cause(s)
Possible Solution(s)
No bG with Bolus
You have not tested your bG and
are attempting to deliver a bolus.
It is recommended to test bG before delivering insulin.
Pairing Failed
The attempt to pair the meter
and pump was unsuccessful.
Restart the pairing process. See the Advanced
Owner’s Booklet.
Pump Not Available The pump is out of the
communication range of the
meter.
• Ensure the meter and pump have
Bluetooth
wireless technology turned on.
• Ensure the meter and pump are within
communication range.
• If the low battery icon is displayed on the meter,
the meter cannot communicate with the pump.
Replace the batteries.