270
Electromagnetic Compatibility
This meter meets the electromagnetic immunity requirements as per EN ISO 15197 Annex A. The chosen
basis for electrostatic discharge immunity testing was basic standard IEC 61000-4-2. In addition, it meets
the electromagnetic emissions requirements as per EN 61326. Its electromagnetic emission is thus low.
Interference from the meter to other electrically driven equipment is not to be anticipated.
Performance Analysis
The performance data for the Accu-Chek Aviva Combo System (Accu-Chek Aviva Combo Meter with Accu-Chek
Aviva Test Strips) were obtained using capillary blood from diabetic patients (method comparison, accuracy),
venous blood (repeatability), and control solution (reproducibility). The system is calibrated with venous blood
containing various levels of glucose. The reference values are obtained using the hexokinase method. For
method comparison, the results were compared with results obtained using the hexokinase method with
deproteinization (automatic analyzer). The hexokinase method is traceable to an NIST standard.
Measuring Principle
Refer to your test strip package insert for more information.
w
WARNING
• Strong electromagnetic fields may interfere with the proper operation of the meter. Do not use this
meter close to sources of strong electromagnetic radiation.
• To avoid electrostatic discharge, do not use the meter in a very dry environment, especially one in
which synthetic materials are present.
53895_00048010161_E.indb 270
9/4/15 9:34 AM
Summary of Contents for Aviva
Page 6: ...iv 53895_00048010161_E indb 4 9 4 15 9 34 AM...
Page 12: ...6 53895_00048010161_E indb 6 9 4 15 9 34 AM...
Page 20: ...14 53895_00048010161_E indb 14 9 4 15 9 34 AM...
Page 52: ...46 53895_00048010161_E indb 46 9 4 15 9 34 AM...
Page 76: ...70 53895_00048010161_E indb 70 9 4 15 9 34 AM...
Page 268: ...262 53895_00048010161_E indb 262 9 4 15 9 34 AM...
Page 282: ...276 53895_00048010161_E indb 276 9 4 15 9 34 AM...
Page 306: ...300 53895_00048010161_E indb 300 9 4 15 9 34 AM...