Display Shows
Possible Cause(s)
Possible Solution(s)
Bolus Delivery
Unavailable
The meter cannot communicate
with the pump.
• Retry or set to Manual Pump.
• Ensure the meter and pump have
Bluetooth
wireless technology turned on.
• Ensure the meter and pump are within
communication range.
• If the low battery icon is displayed on the meter,
the meter cannot communicate with the pump.
Replace the batteries.
The pump is currently in Stop
mode.
Take the pump out of Stop mode.
The pump is currently delivering
a bolus.
Allow the current bolus delivery to complete before
delivering the next bolus.
Bolus Too High
The bolus is above the allowed
amount. The bolus was set to
50 Units.
Check the accuracy of all entries. If necessary,
contact your healthcare professional.
Bolus Too Low
The insulin pump cannot deliver
a 0.1 Unit of multiwave bolus.
The pump will adjust the bolus to 0.2 Units.
Carbohydrate Ratio Carbohydrate ratio is outside of
the acceptable meter range.
Check your entries and contact your healthcare
professional to determine the appropriate settings.
53895_00048010161_E.indb 249
9/4/15 9:34 AM
Summary of Contents for Aviva
Page 6: ...iv 53895_00048010161_E indb 4 9 4 15 9 34 AM...
Page 12: ...6 53895_00048010161_E indb 6 9 4 15 9 34 AM...
Page 20: ...14 53895_00048010161_E indb 14 9 4 15 9 34 AM...
Page 52: ...46 53895_00048010161_E indb 46 9 4 15 9 34 AM...
Page 76: ...70 53895_00048010161_E indb 70 9 4 15 9 34 AM...
Page 268: ...262 53895_00048010161_E indb 262 9 4 15 9 34 AM...
Page 282: ...276 53895_00048010161_E indb 276 9 4 15 9 34 AM...
Page 306: ...300 53895_00048010161_E indb 300 9 4 15 9 34 AM...