Display Shows
Possible Cause(s)
Possible Solution(s)
Beep/Vibrate Off
Beep and Vibrate settings cannot
be turned off at the same time.
Ensure that either the Beep or Vibrate setting is turned
on. See Chapter 5, Changing Meter Settings.
Below Hypo
Warning Limit
Your test result is below the hypo
warning limit set in the meter.
• Treat your low blood glucose as recommended
by your healthcare professional.
• The meter displays a recommended number of
carbohydrates for you to eat, and then retest
your blood glucose.
Bolus Advice
Bolus Advice data
deleted
Bolus advice data integrity
cannot be confirmed.
Wait 8 hours for accurate bolus advice.
IMPORTANT:
• Insulin doses and meals taken before the Bolus Advice warning are no longer
reflected in the bolus advice calculation.
• The meter may not reflect the pump bolus history, however the bolus history is
available on the pump.
Bolus Advice Not
Setup
Advice is not provided unless it
is set up through Bolus Advice in
Settings Menu.
You can continue to use the meter without bolus
advice or see Chapter 5, Changing Meter Settings for
instructions on how to set up bolus advice.
Bolus Advice
Timeout
Bolus advice is no longer
available for this bG result.
Review data that was saved in My Data. Then, decide
to modify data, add data, or begin a new bolus advice
session.
53895_00048010161_E.indb 248
9/4/15 9:34 AM
Summary of Contents for Aviva
Page 6: ...iv 53895_00048010161_E indb 4 9 4 15 9 34 AM...
Page 12: ...6 53895_00048010161_E indb 6 9 4 15 9 34 AM...
Page 20: ...14 53895_00048010161_E indb 14 9 4 15 9 34 AM...
Page 52: ...46 53895_00048010161_E indb 46 9 4 15 9 34 AM...
Page 76: ...70 53895_00048010161_E indb 70 9 4 15 9 34 AM...
Page 268: ...262 53895_00048010161_E indb 262 9 4 15 9 34 AM...
Page 282: ...276 53895_00048010161_E indb 276 9 4 15 9 34 AM...
Page 306: ...300 53895_00048010161_E indb 300 9 4 15 9 34 AM...