Display Shows
Possible Cause(s)
Possible Solution(s)
E-58
Temp. Error
The temperature is above or
below the proper range for the
meter.
Move the meter to an area within the proper
temperature range indicated for test strip use
in the test strip package insert. Wait 5 minutes
before turning on the meter. Repeat the test. Do not
artificially heat or cool the meter.
E-59
Battery Empty
The batteries are extremely low. Insert new batteries. See Chapter 7, Care and
Maintenance.
E-60
Time/Date Error
• The time and date settings
may be incorrect.
• You have changed the
batteries.
Ensure the time and date are correct and adjust, if
necessary. See Chapter 5, Changing Meter Settings.
HI bG Warning
Your blood glucose may be
higher than the measuring range
of the system.
• If you are experiencing any of the common
symptoms of high blood glucose, contact your
healthcare professional immediately.
• Treat your high blood glucose as recommended
by your healthcare professional.
• Consider checking bG, ketones, and insulin.
• See the pump User Guide for additional solutions.
Incorrect Pump
Time/Date
Pump time/date incorrect.
Set time/date on the pump (see pump User Guide).
53895_00048010161_E.indb 255
9/4/15 9:34 AM
Summary of Contents for Aviva
Page 6: ...iv 53895_00048010161_E indb 4 9 4 15 9 34 AM...
Page 12: ...6 53895_00048010161_E indb 6 9 4 15 9 34 AM...
Page 20: ...14 53895_00048010161_E indb 14 9 4 15 9 34 AM...
Page 52: ...46 53895_00048010161_E indb 46 9 4 15 9 34 AM...
Page 76: ...70 53895_00048010161_E indb 70 9 4 15 9 34 AM...
Page 268: ...262 53895_00048010161_E indb 262 9 4 15 9 34 AM...
Page 282: ...276 53895_00048010161_E indb 276 9 4 15 9 34 AM...
Page 306: ...300 53895_00048010161_E indb 300 9 4 15 9 34 AM...