235
6
The errors the meter may display are:
Error Title
Error Message
Left Soft Key Right Soft Key
E-51
Bad Strip Error
Reinsert or replace with new Strip
E-52
Code Key Error
Turn Meter off, reinsert or replace Code Key
1
Menu
2
OK
3
E-53
Bad Test Error
Retest with new Strip
E-54
Not Enough Sample
Retest with new Strip
E-55
Code Key Expired
Turn Meter off, replace Code Key and Strips
1
Menu
2
OK
3
E-56
Sample Applied Early
Retest with new Strip
E-57
Electronic Error
Remove batteries, wait 20 seconds, replace
batteries
E-58
Temp. Error
Move Meter to correct temperature and wait
5 minutes
Menu
4
OK
4