260
Display Shows
Possible Cause(s)
Possible Solution(s)
Test Strips Expiring The test strips expire at the end
of the current month.
• This message may appear when using a white
activation chip in the meter. It means the test
strips expire at the end of the current month. At
the end of the month, discard the white
activation chip and any remaining test strips.
Insert a black activation chip. Make sure the time
and date in the meter are correct. If you need a
black activation chip, please contact Accu-Chek
Customer Care at 1-800-688-4578.
• Ensure the time and date are correct. If the time
or date are not correct, change the time and date
on the pump. See Chapter 5, Changing Meter
Settings.
Time Mismatch
The time or date on the meter
is more than 5 minutes different
than the time or date on the
pump.
The meter time and date were changed to match the
pump time. Ensure the time and date are correct. If
the time or date are not correct, change the time and
date on the pump.
i
NOTE
• For additional information on errors and warnings, see Chapter 6, Icons, Reminders, Warnings, and
Errors.
53895_00048010161_E.indb 260
9/4/15 9:34 AM
Summary of Contents for Aviva
Page 6: ...iv 53895_00048010161_E indb 4 9 4 15 9 34 AM...
Page 12: ...6 53895_00048010161_E indb 6 9 4 15 9 34 AM...
Page 20: ...14 53895_00048010161_E indb 14 9 4 15 9 34 AM...
Page 52: ...46 53895_00048010161_E indb 46 9 4 15 9 34 AM...
Page 76: ...70 53895_00048010161_E indb 70 9 4 15 9 34 AM...
Page 268: ...262 53895_00048010161_E indb 262 9 4 15 9 34 AM...
Page 282: ...276 53895_00048010161_E indb 276 9 4 15 9 34 AM...
Page 306: ...300 53895_00048010161_E indb 300 9 4 15 9 34 AM...