5 Flowmeter Primary: MS2: DN 1.5 to DN 6 [1/16” to 6”]
D184B111U02
FCM2000
21
5.1.1 Supplementary
Installation
Information
In the high temperature design the multi-connection plug is separated from the sensor housing by a pipe.
Thereby the plug can still be accessed even when the sensor is insulated.
The mounting bracket included with the shipment should always be used. The bracket should be secured
to a wall or a steel framework (vibration free and mechanically stable).
Fig. 12:
Connection nipple
Fig. 13:
Installation DN 1.5 – vertical
Connection nipple
Connection
nipple
!
Important!
If there are large differences between the fluid and ambient temperature the sensor must be insulated, to
prevent two phase flow and accuracy effects. This is especially important for low flowrates.
The sensor must
always
be completely filled with homogeneous liquid or a single phase gas, otherwise the
accuracy could be adversely affected.
For air/gas in volatile fluids horizontal installations are recommended.
Fig. 14:
Installation DN 1.5 – horizontal
≥
2 mm
≥
0 mm
Summary of Contents for CoriolisMaster FCM2000
Page 6: ...Contents 6 FCM2000 D184B111U02...
Page 91: ...11 Certificates D184B111U02 FCM2000 91 Fig 43...
Page 92: ...11 Certificates 92 FCM2000 D184B111U02 11 2 EU Type Examination Certificate Fig 44...
Page 93: ...11 Certificates D184B111U02 FCM2000 93 Fig 45...
Page 94: ...11 Certificates 94 FCM2000 D184B111U02 Fig 46...
Page 95: ...11 Certificates D184B111U02 FCM2000 95 Fig 47...
Page 96: ...11 Certificates 96 FCM2000 D184B111U02 Fig 48...
Page 97: ...11 Certificates D184B111U02 FCM2000 97 Fig 49...
Page 98: ...11 Certificates 98 FCM2000 D184B111U02 Fig 50...
Page 99: ...11 Certificates D184B111U02 FCM2000 99 Fig 51...
Page 100: ...11 Certificates 100 FCM2000 D184B111U02 11 3 PED Certificate of Compliance Fig 52...
Page 101: ...11 Certificates D184B111U02 FCM2000 101 11 4 EHEDG Certificate Fig 53...
Page 103: ......