8 Start-Up
82
FCM2000
D184B111U02
Fig. 38:
•
Ignition Type, MC26, MC27
II 1/2G EEx emd [ib] IIC T6 for meter sizes
≥
DN 50 [2”]
II 2G EEx emd [ib] IIC T6 for meter sizes
≥
DN 40 [1-1/2”]
II 2D T115°C...T
fluid
IP67
•
TÜV 99 ATEX 1443X
•
Ambient Temperature (
−
40 °C)
−
20 °C to +60 °C
•
The Temperature Classes, a function of the ambient and fluid temperatures, is listed in the EU-Type Examination Certificate.
•
The supply voltage value for the converter is listed on the Name Plate.
•
The Ignition Type for the output circuits is defined by the external circuits connected. It can be configured as all „intrinsically safe“ or
„non-intrinsically safe“. A combination of „intrinsically safe“ and „non-intrinsically safe“ is not permissible. The test voltage for the
output circuits for both Ignition Types is U
T
= 60 V.
•
The safety specifications for intrinsically safe circuits may be found in the EU-Type Examination Certificate.
•
Care should be exercised to assure that the cover over the supply power connections is correctly installed. For intrinsically safe
output circuits the connection area may be opened.
•
It is recommend that the cable connectors, included with the shipment, used for the output circuits correspond to the Ignition
Type: intrinsically safe
→
light blue; non-intrinsically safe
→
black.
•
The flowmeter primary and the converter housing are to be connected to Potential Equalization.
For intrinsically safe circuits, Potential Equalization is to be maintained along the entire circuit.
•
The interior of the meter pipe for all flowmeter primaries in sizes DN 50 to DN150 [2” to 6”] is classified Category 1 (Zone 0), that for
meter sizes
≤
DN 40 [1-1/2”] only corresponds to Category 2 (Zone 1). The Temperature Classes as a function of the ambient and
maximum fluid temperatures are listed in the table.
•
If the flowmeter primary is to be insulated, the max, insulation thickness is 100 mm.
The converter housing may not be insulated.
•
The corrosion resistance of the meter pipe materials to the fluid must be considered.
Insulation
The pipeline and the flowmeter primary insulation is to be
installed as follows.
The max. insulation thickness is 100 mm.
ma
x. 10
0
mm
31
32
Current output, active
33
34
Current output, passive
51
52
Pulse Output, passive
41
42
Contact output, passive
81
82
Contact input, passive
L
N
85 V - 252 V
1+
2-
24 V AC/DC
Output circuit
Supply voltage
Pressure tight area “d”
Connection Area
Converter
Flowmeter Primary
“ib
“e”
PA
T
Coriolis
F
“m”
PA
PA
„e“
!
Attention!
It is essential that the temperature specifications in the Ex-Approval in Chapter 7.2.2. be observed.
Summary of Contents for CoriolisMaster FCM2000
Page 6: ...Contents 6 FCM2000 D184B111U02...
Page 91: ...11 Certificates D184B111U02 FCM2000 91 Fig 43...
Page 92: ...11 Certificates 92 FCM2000 D184B111U02 11 2 EU Type Examination Certificate Fig 44...
Page 93: ...11 Certificates D184B111U02 FCM2000 93 Fig 45...
Page 94: ...11 Certificates 94 FCM2000 D184B111U02 Fig 46...
Page 95: ...11 Certificates D184B111U02 FCM2000 95 Fig 47...
Page 96: ...11 Certificates 96 FCM2000 D184B111U02 Fig 48...
Page 97: ...11 Certificates D184B111U02 FCM2000 97 Fig 49...
Page 98: ...11 Certificates 98 FCM2000 D184B111U02 Fig 50...
Page 99: ...11 Certificates D184B111U02 FCM2000 99 Fig 51...
Page 100: ...11 Certificates 100 FCM2000 D184B111U02 11 3 PED Certificate of Compliance Fig 52...
Page 101: ...11 Certificates D184B111U02 FCM2000 101 11 4 EHEDG Certificate Fig 53...
Page 103: ......