6.6.15.8
Value of the Alarm Current at Low Alarm
In this menu the value of the Low-Alarm-Current can be entered. The Low-Alarm-Current can be set be-
tween the limits of 2
−
3.6 mA.
6.6.15.9
Value of the Alarm Current at High Alarm
In this menu the value of the High-Alarm-Current can be entered. The High-Alarm-Current can be set be-
tween the limits of 21 mA and 26 mA.
6.6.16 Submenu
Switched
Contacts
In this submenu the function assigned to the contact input and the contact output can be defined
6.6.16.1
Contact input
This menu can be used to define the function of the contact input. The following functions are available:
•
No function
•
Ext. zero return
(Current and pulse outputs are set to 0 % flowrate. Internal totalizers are halted.)
•
Totalizer reset
(resets all mass and volume flow totalizers)
Low Alarm
3.5 mA
High Alarm
23.5 mA
Submenu
Switched contacts
Contact input
Totalizer reset
6 Converter
Summary of Contents for CoriolisMaster FCM2000
Page 6: ...Contents 6 FCM2000 D184B111U02...
Page 91: ...11 Certificates D184B111U02 FCM2000 91 Fig 43...
Page 92: ...11 Certificates 92 FCM2000 D184B111U02 11 2 EU Type Examination Certificate Fig 44...
Page 93: ...11 Certificates D184B111U02 FCM2000 93 Fig 45...
Page 94: ...11 Certificates 94 FCM2000 D184B111U02 Fig 46...
Page 95: ...11 Certificates D184B111U02 FCM2000 95 Fig 47...
Page 96: ...11 Certificates 96 FCM2000 D184B111U02 Fig 48...
Page 97: ...11 Certificates D184B111U02 FCM2000 97 Fig 49...
Page 98: ...11 Certificates 98 FCM2000 D184B111U02 Fig 50...
Page 99: ...11 Certificates D184B111U02 FCM2000 99 Fig 51...
Page 100: ...11 Certificates 100 FCM2000 D184B111U02 11 3 PED Certificate of Compliance Fig 52...
Page 101: ...11 Certificates D184B111U02 FCM2000 101 11 4 EHEDG Certificate Fig 53...
Page 103: ......