AD-4405A Weighing Indicator
Page
59
Command Format
Description of Commands
When performing a command, the received command or replay data is sent
back.
When the received command can not be performed such as in the "busy" state,
the code "
I
" is sent back. In this case, insert a delay time. Electrical noise may
be the cause of this error.
When receiving an undefined command (unknown command), a "?" is sent back.
When the memory is insufficient to store the data of the UFC commands, an “M”
is sent back.
Optional addresses can be added to a command. The address form is
"@address" and the address is specified at F-Function
f06
. The reply (data or
error code) is also sent with the address.
Example: The command is "Display net data". The address is 23.
Sending a command
Sending a command with address
Unperformed command reply
Unknown command reply
Memory overflow (for UFC commands)
16.3. Command Format
M
Terminators
Memory overflow error
Carriage return(0Dh)
Line feed(0Ah)
I
Unperformed command
Terminators
N
M
Terminators
Command
Terminators
Command
N
M
@ 2 3
Address number
Unknown command
Terminators
?
@ 2 3
Address number
16.3.
.