
Page
12
AD-4405A Weighing Indicator
CAL
The key to store all of the parameters into memory and display
Caloff
.
Press the
ON/OFF
key to turn off the display.
Note that the
ON/OFF
key does not function alone in the calibration mode.
When a value is mistakenly set, press the ESC key while holding the
ON/OFF key to finish the calibration mode without storing the value. After
CanCel
is displayed, press the ON/OFF key to exit from the calibration mode
and turns the display off.
NOTE:
Caloff
can be displayed by pressing the
+/-
key while pressing the
ON/OFF
key,
instead of
CAL
key.
Caution
The maximum display is less than or equal to 40000 divisions. This number
is calculated from the maximum capacity divided by the minimum division.
Check the accuracy of the weighing instrument periodically.
Recommended mass, use a mass heavier than 2/3 maximum capacity.
Calibrate the scale, if it is moved to another location or the environment
has changed.
It is not necessary to set the gravity acceleration correction, when calibrating
the scale with a calibration mass at the place where the scale is used.
Enter the stable weighing data while the STABLE mark is displayed. If
unstable data is used, it may cause a weighing error. Arrange the
condition using the F00 filter function.
The span calibration needs the zero calibration data. We recommend
that you perform the span calibration immediately after the zero
calibration.
If you use the dual range function of the multi-interval scale, perform the
"Range Function", "Zero Calibration" and "Span Calibration".
5.2. Calibration Procedure
5.2.1. Configuring a Weighing Instrument
This section explains how to set capacity, resolution, decimal point position and display
format, weighing range and unit. Perform this procedure when installing the indicator.
When
Cal5et
is displayed, enter the setting mode by pressing the ENTER key.
Setting the range and unit.
Single Range
Select the resolution, decimal point position and format.
Specify the weighing capacity.
5. Calibration