![A&D AD-4405A Instruction Manual Download Page 26](http://html1.mh-extra.com/html/aandd/ad-4405a/ad-4405a_instruction-manual_3091937026.webp)
Page
24
AD-4405A Weighing Indicator
Display and Other General Functions
Item
Parameter
Description
*
0
5 times/s
f04
Display update rate
1
10 times/s
The selection in the unstable
condition.
1
x
Key click (ON/OFF)
2
x
LoLo / Zero band
3
x
LO
4
x
OK
5
x
HI
f05
[Type2]
Buzzer
6
x
HiHi /
Batch finish / Full
Left: Item, select using the
ZERO
key
Right: Parameter, select using the
0
-
5
keys
0: OFF
1: Continuous
2: 4 times/s
3: 2 times/s
4: 1 time/s
5: 1 time/2s
Initial value is
11
, others x
0
.
f06
Device ID (Address)
00
to
99
Command address
or Device ID
Initial setting is
00
.
*
0
Disable
f07
Counting function
1
Enable
*: Initial
settings.
s: second.
Key Switch
Item
Parameter
Description
2 0
f12
[Type2]
Disabling key
switches
Left: Key
Right:
0
Enable
1
Disable
ZERO
key to select a key
0,
1
key to select
Enable/Disable
When
f13
,
f14
and
f15
are set to
17
and the terminal is connected
to the common, all keys are
enabled, regardless of the
f12
setting.
*: Initial
settings.
*
0
Enable
1
Disable
1
CODE key
2
SAMPLE key
3
SETPOINT key
4
PT key
5
TOTAL key
6
M+ key
7
START key
8
STOP key
9
FEED key
a
MODE key
b
ZERO key
C
TARE key
d
NET/GROSS key
e
PRINT key
f
ZERO+TARE = CAL
6.2. F-Functions