AD-4405A Weighing Indicator
Page
35
8.2. Display and Operation
Action of Accumulation
When accumulating data, the display blinks once.
If the accumulated data is stored, the M+ mark is displayed.
Caution
This function can not accumulate data with a different unit. Specify a unit
before use.
Display of Accumulated Data
With
Cf08
=
1
(Effective accumulation function), press the
TOTAL
key. After
total
appears,
the total data is displayed with the M+ mark blinking.
Pressing the
TOTAL
key alternatively displays the accumulated data and the
accumulation count. Press the
ESC
key to return to the weighing mode.
The total data can be output. Refer to "Output of Accumulated Data".
Undoing the Accumulated Data
The last weight data can be deduced from the accumulated data unless new data
has been accumulated.
Step 1 Press the
TOTAL
key to display
total
and the accumulated data.
Step 2 Press and hold the
+/-
key for more than 3 seconds. The display blinks once and
the data accumulated before accumulating the last weight data is displayed.
Caution
External input can not be used.
Clearing the Accumulated Data
Step 1 Press the
TOTAL
key to display
total
and the accumulated data.
Step 2 Press and hold the
ZERO
key for more than 3 seconds. The display blinks once
and the accumulated data is cleared.
Caution
External input can not be used.
Initializing the Data Number and Clearing the Accumulated Data at the same time
When the data number is included with the data of the serial data output or the
printer, initializing the data number and clearing the accumulated data can be
done at same time.
If using both serial output and printer data number, all data numbers are
initialized by this operation.
Step 1 Press
TOTAL
key to display
total
and the accumulated data.
Step 2 Press and hold the
ZERO
and
+/-
keys at the same time for more than 3 seconds.
The display blinks once and the accumulated data is cleared. And the data
number is initialized (1).
Caution
External input can not be used.
8.
Accumulation