AD-4405A Weighing Indicator
Page
55
Description of "Command Mode" (For Serial Interface)
Function
f40
=
5
Operation
The serial data can be output by the serial interface command only.
When multiple instruments are used, using this mode will avoid data conflicts
or accidental data output by mis-operation.
Description of "Delay for Continual Data"
Function
f38
,
f42
Operation
This function can be used in the "Auto print mode" and "Manual print mode".
When using a non-buffered printer, set
f42
to
3
and
f43
to
1
(or
f38
to
1
).
15.2. Data
Number
The data number increments automatically with each serial data output or print out
respectively from 1 to 99999 (1 comes after 99999). The starting number can be set
using the keys.
Setting the Serial Data Number
Press the
CODE(1)
key while pressing the
key at the weight display. The display
changes to
data r
then
r
xxx. “xxx” is the next data number. Use the
0
-
9
keys to enter
the number, then press the
ENTER
key to store the number and return to the weight
display.
Setting the Print Data Number
Press the
SAMPLE(2)
key while pressing the
key at the weight display. The
display changes to
data p
then
p
xxx. “xxx” is the next data number. Use the
0
-
9
keys
to enter the number, then press the
ENTER
key to store the number and return to the
weight display.
Note
1. When setting the number, the following keys are available.
+/- key
Sets the number to 1.
ZERO
key Sets the number to 0.
ESC
key
Returns to the weight display without changing the number.
2. When the number is set to 0, it is entered as 1 automatically.
15.3. Interval Data Output / Print
The AD-4405A can output or print data with a constant time interval using the
internal calendar/clock function. The time intervals for serial data output and for the
printer can be set individually from 1 second up to 23 hours and 59 minutes.
Caution The interval time must be set to allow enough time to finish output or
printing out.
15. Common Items for Printer and Data Output