AD-4405A Weighing Indicator
Page
47
12. Counting Function
This function determines the number of objects in a sample based on the unit
weight. The unit weight is stored in one of the code memory data in non-volatile
memory, and is not lost even if power is turned off.
12.1. Using the Counting Function
Preparation:
Set
f07
to
1
in the F-Functions.
Select a code memory number. Refer to “9.1. Using Code Memory”.
Register the unit weight. Refer to “12.2. Unit Weight Registration”.
Press
the
MODE
key to enter the counting mode. The unit in the display changes to
pcs
.
In the counting mode, TARE, ZERO, NET/GROSS and other operations are the same
as that in the normal weighing mode.
12.2. Unit Weight Registration
There are three methods to register the unit weight. By weighing the actual samples to
determine the unit weight, by digital input of the unit weight, or by command via the
serial interface.
The unit weight is one item of the code memory data. Select the code memory
number and start the unit weight registration in the weight display or piece count
display. The unit weight registration can not be done from the code memory
number display.
There is no counting accuracy improvement function by re-calculating the unit
weight based on the actual weight.
Weigh Actual Samples
When the unit weight is unknown, prepare some samples to determine the unit weight.
Step 1 Enter the counting mode.
Step 2 Press the
SAMPLE
key to enter the unit weight registration mode.
5 0
or
5 -
is displayed and unit
pcs
blinks. The left side of the display is the number of
samples. The right side of the display indicates the weight,
0
means that the
weight is zero and
-
means that the display is not at zero.
Step 3 Select the number of samples and place the samples specified on the weighing
unit. Use the following keys:
0
-
9
key .................The key to set the number of samples. Maximum is 9999.
+/-
key.....................The key to select the number of samples, 5, 10, 20, 50,100.
ZERO
key ...............The key to set the zero point.
TARE
key................The key to tare (when using a container).
12. Counting Function