AD-4405A Weighing Indicator
Specifying the Weighing Range of the First Range
Step 3 After displaying
Cap
for 2 seconds, the first range or the weighing capacity of single
range will be displayed. When dual range is used,
Cap1
is displayed for 2 seconds.
Triangle 1 will be displayed.
Specify the parameter using the
0
-
9
keys and press the
ENTER
key to store it
and proceed to the next step. When pressing the
ESC
key, regardless of what is
displayed, the indicator will proceed to the next step without changing the
parameter. The next step is Zero Calibration in single range or the second range
resolution in dual range.
Specifying the Second Range Resolution
Step 4 After displaying
range2
for 2 seconds, the resolution with decimal point and triangle
2 will be displayed. Specify the second range resolution in the same way as the
first range. The decimal point cannot be moved. Specify the second range resolution
greater than the first range.
Press the
ENTER
key to store the parameter and proceed to the next step. When
pressing the
ESC
key, regardless of what is displayed, the indicator will proceed to
the next step without changing the parameter.
Specifying the Second Range Capacity
Step 5 After displaying
Cap2
for 2 seconds, the capacity with unit and decimal point is
displayed. Specify the capacity in the same way as the first range capacity. The
value should be greater than the first range capacity.
Press the
ENTER
key to store the parameter and proceed to Zero Calibration. When
pressing the
ESC
key, regardless of what is displayed, the indicator will proceed to
the next step without changing the parameter.
5.2.2. To Get Stabilized Data
Step 6 Maintain the following conditions to calibrate the scale (indicator) correctly.
Maintain a constant temperature, stable power and stable input voltage from the
load cell.
Avoid direct sunshine or the near the outlet of an air conditioner.
Do not install the scale (indicator) where there is a strong magnetic field.
Step 7 Turn the display on and leave it for several minutes.