5 Scanning modes and settings
Click the
button to display barcode label area. To rotate the label area by 180°, click
.
Figure 31 – Barcode label area window
5.1.2.2 Focus Indicator
Click
to activate the
Focus indicator
function. Search for the sharpest focus on the current field
of view automatically by clicking
-
Auto focus
button.
To define focus limits, select
Limit the range
. Set the upper and lower limits by selecting the desired
value in the selection boxes. By using
Set +-Range
, the
Lower limit
value is set to -32 of the current
focus value and the
Upper limit
value is set to +32 of the current focus value. This is useful if
Auto
Focus
does not find the focus on the specimen, but on the coverslip. Also, this makes the focusing
sequence faster.
For manual focusing define the manual focus step size by adjusting the
Step size
slider and use the
two arrow buttons to focus manually. This is useful when you are interested in an area that remained
blurry with
Auto Focus
(for example, the specimen is thicker in that area). This does not influence the
way slides are digitized in Auto focus mode.
Figure 32 – Focus indicator panel
5.1.2.3 Exposure Indicator
Activate Exposure indicator function by clicking
. Click
to calculate exposure time for the
actual Live image. In Brightfield mode there is no option for manually setting the values. Exposure
time can be displayed in microsecond, millisecond, and also in second.
Figure 33 – Window of Exposure Indicator in Brightfield scan mode