Pannoramic SCAN II 2.0 User’s Guide
6 Troubleshooting and Maintenance
Virtual slide is improperly focused almost across its entire surface
•
Focus range enabled
•
Disable
Limiting focus range
.
Poor image quality
•
Imaging error (for example, chromatic aberrations)
•
Contact Thermo Fisher ScientificService for support.
Warning!
Do not modify the camera driver settings, as the default factory
settings give the most accurate images.
Virtual slides show poor stitching in Pannoramic Viewer
•
Scan camera not properly adjusted (for example, due to loose fixture)
•
Contact Thermo Fisher Scientific Service for support.
Slide is often positioned incorrectly
•
Guiding fault in transport system
•
Check if faulty positioning disappears after the selection of a different type of
magazine (for example, universal magazine).
•
Contact Thermo Fisher Scientific Service for support.
Digitization process is extremely slow
•
Extended focusing
option is activated
•
Deactivate the
Extended focusing
option if not required.
•
Z-stack
option is activated
•
Deactivate the
Z-stack
option if not required.
Image is out of focus (in Manual mode)
•
Focus range limit is activated
•
Deactivate the
Limiting focus range
option.
June 4, 2018 - Rev. 2
3DHISTECH Ltd.
122(142)
For research and education uses only, not for use in diagnostic procedures.