Pannoramic SCAN II 2.0 User’s Guide
5 Scanning modes and settings
•
Thresholding:
The manually specified value is used for thresholding during scanning. Preview
image of the loaded slide is displayed in the Preview window, and when modifying any of the
parameter values, it will have effect on the preview image. If the threshold value presented
on the Microscope Controller preview window is modified, this value is automatically applied
by the profile in the Profile Settings panel.
◦
Specimen threshold
This value is definable manually:
This sets the sensitivity of the specimen detection. To
set the desired threshold value manually, change the setting in the
Specimen threshold
level
selection box in the range of 0-255. The default value is 220. Smaller values result in
a more sensitive scanning, meaning that more areas will be digitized, however, this may
also include specks or other imperfections of the slide. Greater values result in a less
sensitive scanning, but it is possible that some of the smaller specimen areas will not be
digitized.
June 4, 2018 - Rev. 2
3DHISTECH Ltd.
48(142)
For research and education uses only, not for use in diagnostic procedures.