Pannoramic SCAN II 2.0 User’s Guide
3 Operating Slides and Magazines
3 Operating Slides and Magazines
3.1 Preparing Slides and Magazines
This section demonstrates how to prepare slides and magazines. You can use slides and coverslips
that meet the following specifications:
Slides
Cover slips
Length
75.0 to 76.0 mm
max. 50 mm
Width
25.0 to 26.0 mm
max. 24 mm (recommended: 22 mm)
Thickness
0.9 to 1.2 mm
No. 1 and No. 1.5 (0.13 to 0.16 mm and 0.16 to 0.19 mm)
Corners
45° beveled corners
Edges
Grounded or cut
Warning!
Always wear personal protective equipment (safety goggles, gloves and
coat) when working with biohazardous material or slides.
Warning!
Do not use a slide when it is broken or its edges are not properly cut or
grounded. Using such slides can lead to injury.
Warning!
Ensure that no part of the mounted coverslip protrudes from the edges
of the slide. Use slides only with completely dried embedding medium.
If not so, the coverslip can be shifted that makes slide loading less
reliable. Not completely dried embedding might also influence the
operation of objectives and other device parts.
Caution!
Do
NOT
scratch any circular marker onto the glass surface of a slide,
use a soft tip pen instead for marking area of interest.
June 4, 2018 - Rev. 2