Pannoramic SCAN II 2.0 User’s Guide
6 Troubleshooting and Maintenance
Lumencor SOLA SM II illumination unit
•
Keep inflammable and combustible materials outside the range of
Lumencor SOLA SM II
illumination unit.
•
Do not expose yourself to the UV radiation of the
Lumencor SOLA SM II
illumination unit.
•
When the
Lumencor SOLA SM II
illumination unit is switched on:
◦
NEVER
look directly into the fiber-optic cable.
◦
NEVER
remove the fiber-optic cable from the device.
◦
NEVER
connect the fiber-optic cable to the device.
◦
NEVER
remove the housing from the device.
6.1.1 Cleaning the device
•
Before cleaning, switch off the device and pull the plug out.
•
Do not let liquid penetrate into the system.
•
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with
water.
•
To clean greasy surfaces, it is recommended to apply
soapy water
on a piece of lint-free
cotton cloth and gently rub the surface.
6.1.2 Cleaning the magazine
Do not clean the magazines with xylene, because it might harm the plastic. We recommend to apply
Isopropanol
on a piece of lint-free cotton cloth and gently rub the surface instead. In case the
magazine is damaged, use either order a new one, as inserting a deteriorated magazine could lead
into unsafe loading of slides.
6.1.3 Preventive maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storage temperature
above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
•
Maximum operating temperature is 35°C (95°F) with a relative humidity of less than 75%.
•
Install it in a room with proper air ventilation.
•
Do not let dust or fingerprints remain on optical surfaces.
June 4, 2018 - Rev. 2
3DHISTECH Ltd.
120(142)
For research and education uses only, not for use in diagnostic procedures.