Page 66
MAINTENANCE
SYMPTOM POSSIBLE
CAUSES
Backup battery not charging.
•
Blown internal battery fuse.
•
Battery not properly plugged in.
•
Unit not plugged into wall outlet (verify the green AC indicator is
illuminated).
•
~ AC Power Mains harness not properly plugged in (verify the
green AC indicator is illuminated).
•
Unit was not permitted to charge for at least 10 hours.
•
Defective battery.
•
Defective battery charging circuitry.
•
Poor battery health.
SYSTEM FAILURE DETECTED alarm.
•
Multiple possible causes. See Alarm Conditions section for
solutions.
Unit cannot be set to STANDBY.
•
ON/STANDBY Membrane not properly plugged in.
•
Pressure sensed in the Main or Second Cuff (unit will alarm
“CUFF NOT DEFLATED”)
•
ON/STANDBY not fully pressed.
•
ON/STANDBY button defective.
Unit does not turn ON (blank screen).
•
Membrane panel not properly plugged in.
•
ON/STANDBY button defective.
•
Blown Fuse(s).
•
Unit not plugged in and battery fully depleted
•
Unit not plugged in and battery defective or battery fuse blown.
CAUTION - HIGH VOLTAGE ELECTRICAL HAZARD:
High voltage will be present on the power input
module and control board. All service work must be completed by Zimmer or a qualifi ed biomedical
engineer or other person thoroughly familiar with electronic medical devices within the healthcare
organization.
REPLACEMENT PARTS
The following is a list of fi eld replacement parts that can be ordered from Zimmer:
PARTS LIST
PART NUMBER
DESCRIPTION
62800000130
Replacement Fuses (5-Pack)
62800011380
Replacement Fuse Drawers (5-Pack)
60400011000
Calibration Hose Assembly Kit
60400070200
Replacement Battery
62120080019
Power Cord Clip (5-Pack)
60400021317
Replacement Battery Door
60400090773
Battery Door #8-18 Screw (20-Pack)
60236000100
Pole Clamp Knob
62800011340
ATS Foot Pads (20-Pack)
60401800100
Dual Cuff Hose Red