Page 10
INSTALLATION AND OPERATING INSTRUCTIONS
Button/Icon Title
What it means
Brightness Icon
Appears in the system Settings when adjusting the backlight
brightness.
LOP Button
Controls the Limb Occlusion Pressure (LOP) feature. When the
pulse sensor is in place and cuff applied, pressing this button will
start the process to measure the patient’s LOP and give the user a
recommended Tourniquet pressure (RTP).
Cuff Progress Icon
Indicates that the Tourniquet system is currently testing the
Tourniquet cuff attached to the Main Cuff Ports.
Cuff Success Icon
Indicates that the Tourniquet system has tested the Tourniquet cuff
attached to the Main Cuff Ports and that it passed the leak test.
Cuff Fail Icon
Indicates that the Tourniquet system has tested the Tourniquet cuff
attached to the Main Cuff Ports and that it failed the leak test.
Check Cuff Icon
Indicates that the Tourniquet system has detected a connection or
infl ation problem with the Tourniquet cuff attached to the Main Cuff
Ports when attempting to leak-test the cuff.
INFLATE Button
(INFL)
Controls infl ation of the respective cuff. The Main Cuff and Second
Cuff have separate INFLATE buttons.
DEFLATE Button
(DEFL)
Controls defl ation of the respective cuff. The Main Cuff and Second
Cuff have separate DEFLATE buttons.
Okay/Success Button
Allows user to approve, save and exit specifi c functions.
Cancel/Fail Button
Allows user to exit specifi c functions without saving.
Retry/Reload/Reset
Button
Allows user to restart, reset, or retry specifi c functions.
Next Button
Allows user to progress to the next step in specifi c functions.
Progress Icon
Appears and animates to indicate that a specifi c function is loading
or working.
A.T.S.
Progress Icon
Indicates that a Tourniquet instrument level function is loading or
working.
A.T.S.
Fail Icon
Indicates that a Tourniquet instrument test has failed to meet the
requirements.
A.T.S.
Success Icon
Indicates that a Tourniquet instrument test has successfully met the
requirements.
A.T.S.
w/ Cuff Icon
Appears and
animates
to indicate that the Advanced Leak Detection
feature has detected a potential leak with a cuff, hoses or the
pneumatic connection between the cuff and device during cuff
infl ation.
Pressure Icon
Represents a Pressure function.
Time Icon
Represents a Time function.