Page 43
INSTALLATION AND OPERATING INSTRUCTIONS
Warning
Message
Indicator
Light
Priority
Tone
Condition, Remarks and Appropriate Action
BATTERY IS LOW
Constant
Yellow
Low
Condition: Battery is low (low).
Battery charge is below 30% capacity. Plug unit into ~ AC Power Mains. If the unit is
not plugged in, a battery fail condition will occur and the unit will shut down in a fail
safe mode closing all cuff valves.
While running with a Low Battery Voltage Alarm
Condition, other alarm conditions cannot be guaranteed.
BATTERY IS LOW
Flashing
Yellow
Medium
Condition: Battery is low (medium).
Battery charge is below 20% capacity. Plug unit into ~ AC Power Mains. If the unit is
not plugged in, a battery fail condition will occur and the unit will shut down in a fail
safe mode closing all cuff valves.
While running with a Low Battery Voltage Alarm
Condition, other alarm conditions cannot be guaranteed.
CUFF NOT
DEFLATED
Flashing
Yellow
Medium
Condition: Power switch was activated with cuff pressure.
Cuff pressure greater than zero when ON/STANDBY button pressed. The system
assumes that a procedure is in progress and prevents activation of STANDBY mode.
CUFF INFLATED
Flashing
Yellow
Medium
Condition: A cuff was found pressurized during power up.
Cuff pressure is 50 mmHg or greater at system power-up. The system assumes that
a procedure is in progress and adopts the sensed pressure as the new set-point. The
operator should immediately check the set value to determine if it needs reset.
CUFF INFLATED
Constant
Yellow
Low
Condition: Cuff pressurization issues.
Cuff with non-zero pressure is connected to cuff ports. Defl ate cuff and try again.
CUFF NOT
DEFLATED
Flashing
Red
High
Condition: Cuff not fully defl ated as expected.
Defl ated cuff pressure is greater than 15 mmHg. Check for kinks in hose. If alarm
persists, disconnect hose from cuff. If attempting to set the unit to STANDBY, ensure that
cuff is
fully
defl ated.
CUFF LEAK
Flashing
Red
High
Condition: Cuff or hose is leaking.
Defl ated Cuff fails to infl ate to set-point in at least 30 seconds or Infl ated cuff leaks more
than 15 mmHg in at least 9 seconds. Check all lines and connections.
CUFF DEFLATED
Flashing
Yellow
Medium
Condition: Cuff is defl ated reminder.
Alerts user that the other cuff has been defl ated. When both cuffs are infl ated for a
Bier Block (IVRA), the system remembers and when one cuff is defl ated, warns the
user when the other cuff’s Defl ate (DEFL) button is touched to help avoid accidental
bolus release. A message displaying
CONFIRM DEFLATE?
also appears requesting
confi rmation of fi nal cuff defl ation.
CHECK CUFF
Flashing
Yellow
Medium
Condition: Check the cuff.
Cuff has not begun to infl ate within 1 second of touching the cuff Infl ate (INFL) button.
Check all lines and connections.
LINE OCCLUSION
Flashing
Red
High
Condition: Line is occluded.
An occlusion is present in the cuff tubing. Check for hose kinks or other defects.
CONNECT LOP
SENSOR
Constant
Yellow
Low
Condition: Missing or disconnected LOP sensor detected.
No LOP pulse sensor detected during LOP. Plug LOP pulse sensor in. Reattach LOP
pulse sensor and retry. Replace LOP pulse sensor with known good working sensor and
retry. If problem persists, service the unit.
CHECK LOP
SENSOR
Constant
yellow
Low
Condition: LOP sensor not applied properly or defective.
LOP pulse sensor not properly secured to patient. Check LOP pulse sensor and retry.
Replace LOP pulse sensor with known good working sensor and retry. If problem
persists, service the unit.
LOW SIGNAL
ADJUST SENSOR
Constant
Yellow
Low
Condition: Low signal measured at the LOP sensor.
The LOP signal is too weak. Check LOP pulse sensor and retry. Replace LOP pulse
sensor with known good working sensor and retry. If problem persists, service the unit.