Page 12
INSTALLATION AND OPERATING INSTRUCTIONS
Feature
What it does
1
Carrying Handle
Handle for carrying or holding the Tourniquet system.
2
Touch Screen GUI
Touch-screen based graphical user interface (GUI) for interacting with and
controlling most of the Tourniquet system functions.
3
Power ON/STANDBY
Button
Turns the unit ON or sets the unit to STANDBY.
The illuminated green AC INDICATOR indicates that the unit is plugged
in and is being powered by ~ AC Power Mains. This is the normal means
of operation (battery power is only intended for emergency power loss or
patient transport).
CAUTION:
Ensure both cuffs are fully defl ated and have been
removed from the patient as well as all underlying bandages or
protective sleeve prior to setting the unit to STANDBY.
CAUTION:
•
This button will not set the unit to STANDBY when the cuff
pressure is at a non-zero value.
•
During STANDBY, the power to the
A.T.S.
4000TS instrument
and all instrument functions (i.e. infl ation, defl ation, etc.)
are OFF but power continues to supply the battery charging
circuitry anytime ~ AC Power Mains is present.
4
Alarm Silence Button
Will silence most audible alarms for 30 seconds after the button is
pressed.
An alarm silenced icon will appear in the cuff status display to indicate the
alarm is silenced. Alarms will silence after the full priority tone has played.
CAUTION:
•
When an alarm sounds because of an internal hardware
malfunction, the alarm cannot be silenced.
•
The alarm messages will continue to fl ash on the displays
until the alarm condition is corrected.
4
10
9
12
13
14
11
15
6
7
5
1
17
2
3
16
8