Page 60
MAINTENANCE
CAUTION:
This test will check internal pneumatic connections inside the ATS machine. This test is not
intended to check cuffs or external connections. A leaking external connection unrelated to the ATS
machine will potentially cause this test to show a false leak.
UNIT INFORMATION ACCESS
The system contains an information block that will allow a qualifi ed biomedical engineer or other person thoroughly
familiar with electronic medical devices within the healthcare organization the ability to check on key system elements
that help track usage. This function is located in the SERVICE menu. To access this information perform the following
steps:
1
Access the
Service
menu as described in the
Service Menu Access
section above.
2
Touch the
Unit Info
button to access the Unit Information function interface.
Once opened the system will display information about the system that may be helpful for periodic maintenance.
UNIT INFO
What it indicates
1
Software Version
Displays the current version of software installed.
2
Total Cuff Runtime
Total time in minutes a cuff has been infl ated. This timer counts when either cuff is
infl ated.
3
Total Successful
Calibrations
Total time the system has been successfully calibrated in its life.
4
Total Time Used in
Backup Battery
Total time in minutes the system has been used in backup battery mode.
5
Battery Health
Health state of the battery expressed in percent (%).
6
Last 3 Error Codes
Displays the last error codes encountered. This may initially include error codes
that was intentionally generated during the factory testing.
CAUTION:
If the button combination was not correctly touched and held, the Systems Function Menu will
not allow access and automatically close the menu due to inactivity. While in the Service mode, the system
will keep the function running without automatically closing.
BURN MODE
The system contains a feature that will allow a qualifi ed biomedical engineer or other person thoroughly familiar with
electronic medical devices within the healthcare organization the ability to run the system in a Burn Mode. The Burn
Mode will continuously cycle the pneumatic system by automatically changing pressure (infl ating and defl ating) as
a means to exercise the functions and stress the pneumatic circuit. The function is only intended to be used after
signifi cant service that has repaired or replaced parts of the control circuit, pneumatic circuit or power system. This
function is located in the SERVICE menu. To access this information perform the following steps:
1
2
3
4
5
6
Содержание A.T.S. 4000TS
Страница 1: ...A T S 4000TS Operator Service Manual REF 60 4000 101 00 ENGLISH ...
Страница 75: ......