Page 48
INSTALLATION AND OPERATING INSTRUCTIONS
Error Code
Warning
Message
Indicator
Light
Priority
Tone
Explanation / Appropriate Actions
E0038 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
User interface failure detected.
Specifi cation:
Internal system diagnostics feature has detected a
backlight error. The unit has shut down in a safe state.
Explanation/Action:
The LCD Backlight has failed to function within
specifi cations. Cycle the ON/STANDBY button. If problem persists, note
error code and contact manufacturer.
E0039 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
User interface failure detected.
Specifi cation:
Internal system diagnostics feature has detected a LCD
error. The unit has shut down in a safe state.
Explanation/Action:
The LCD controller is unresponsive. Cycle the
ON/STANDBY button. If problem persists, note error code and contact
manufacturer.
E0040 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
User interface failure detected.
Specifi cation:
Internal system diagnostics feature has detected a Touch
Screen Controller error. The unit has shut down in a safe state.
Explanation/Action:
The Touch Screen controller is unresponsive.
Cycle the ON/STANDBY button. If problem persists, note error code and
contact manufacturer.
E0041 SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Illegal program execution event or data values detected.
Specifi cation:
Internal system diagnostics feature has detected an
illegal program execution event or illegal data. The unit has shut down in
a safe state.
Explanation/Action:
The system has detected an occurrence of an
illegal program execution event or data values. If problem persists, note
error code and contact manufacturer.
1
Some Error Codes are followed by a four-digit numeric code. This code represents detailed information related to the failure. Contact the manufacturer
for information concerning those specifi c codes.
ELECTROMAGNETIC COMPATIBILITY (EMC) GUIDANCE TABLES
The following tables provide guidance on needs and installation of the
A.T.S.
4000TS regarding electromagnetic
compatibility.
EMC GUIDANCE AND DECLARATION - EM EMISSIONS
Guidance and manufacturer’s declaration – electromagnetic emissions.
The
A.T.S.
4000TS is intended for use in the electromagnetic environments specifi ed below. The customer or the user of the
A.T.S.
4000TS
should assure that it is used in such an environment.
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions
CISPR 11
Group 1
The
A.T.S.
4000TS uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The
A.T.S.
4000TS is suitable for use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fl uctuations/
Flicker emissions
IEC 61000-3-3
Complies
Содержание A.T.S. 4000TS