Page 13
INSTALLATION AND OPERATING INSTRUCTIONS
Feature
What it does
5
Main Cuff Ports
Connector ports used to connect the Tourniquet system to the Main
Cuff hoses. The Main Cuff Ports are controlled by the left side
RED
GUI
controls.
CAUTION:
The
A.T.S.
4000TS is designed and tested for use with
Zimmer dual port cuffs and hoses. Zimmer does not recommend
the use of any cuff other than Zimmer dual port cuffs and hoses.
Do not use single port cuffs with the
A.T.S.
4000TS.
6
Second Cuff Ports
Connector ports used to connect the Tourniquet system to the Second
Cuff hoses. The Second Cuff Ports are controlled by the right side
BLUE
GUI controls.
CAUTION:
The
A.T.S.
4000TS is designed and tested for use with
Zimmer dual port cuffs and hoses. Zimmer does not recommend
the use of any cuff other than Zimmer dual port cuffs and hoses.
Do not use single port cuffs with the
A.T.S.
4000TS.
7
Pole Clamp
The Pole clamp is used to mount the unit on an I.V. pole. See
Specifi cations
section
above for details.
CAUTION:
Do not hang articles on the Tourniquet pole that are
not related to Tourniquet use. For stability reasons, do not use an
I.V. Pole with a base less than
≥
27.56 inches (70 cm) in diameter.
8
Hose Hangers
The
A.T.S.
4000TS is equipped with hose hangers at the back of its case.
The cuff hoses can be temporarily hung on the hangers for transport or
when disconnecting from the cuff.
CAUTION:
•
Cuff hoses or the LOP sensor should be the only items to
utilize the hose hangers.
•
Do not hang articles on the Tourniquet’s hose hangers that
may cause the Tourniquet to become unstable.
9
Power Cord
(not shown)
Connects to IEC320 appliance inlet on the Tourniquet system. See
Specifi cations
section above for details.
10
~ AC Power Mains Fuse
Block
Contains ~ AC Power Mains Fuses. Fuse drawer is a 2-pole holder for
5x20mm fuses.
11
Potential Equalization
Conductor Stud
Conductor stud that can be used to provide a connection between the
A.T.S.
4000TS and the potential equalization bus-bar of the electrical
installation.
12
USB Port A
Host port for Zimmer-authorized servicing of the
A.T.S.
4000TS.
CAUTION:
Do not touch the ports or USB cable conductors and
the patient at the same time.
13
USB Port B
Device port for Zimmer-authorized servicing of the
A.T.S.
4000TS.
CAUTION:
Do not touch the ports or USB cable conductors and
the patient at the same time.