INSTALLATION AND OPERATING INSTRUCTIONS
•
The animated Progress icon
is replaced by the text “DIAGNOSTIC OKAY.”
•
“CALIBRATION” and an animated Progress icon
are displayed during the calibration check. Once
complete, “CALIBRATION OKAY” replaces the previous text and icon.
CAUTION:
The
A.T.S.
4000TS will automatically perform an internal Diagnostic and Calibration check when
powered ON. Setting the system to STANDBY and powering back ON between cases will allow the system
to re-perform the automatic Diagnostic and Calibration check.
•
After the startup routine is complete, the default Graphical User Interface (GUI) appears:
•
“
0
” is indicated in the PRESSURE and TIME tiles of both the Main and Second displays.
•
The default pressure is shown on both PRESSURE tiles. The default pressure is pre-set to 250 mmHg. This
setting may be changed as desired.
•
The default time is shown on both TIME tiles. The default time is pre-set to 60 minutes. This setting may be
changed as desired.
CAUTION:
•
If “CALIBRATION ERROR” is displayed, refer to the Calibration Error at Power-On section with the
Maintenance section.
•
If a number other than zero is indicated in the PRESSURE displays, the unit should be calibrated by a
qualifi ed biomedical engineer or other person thoroughly familiar with electronic medical devices via
the Calibration section within the Maintenance section.
MANUAL TESTS AND CHECKS
These manual tests and checks verify certain System Functions and include Pressure and Time set-point tests and
Calibration and Low Pressure Alarm checks.
1
PRESSURE set-point system test
•
Touch the red Main Cuff PRESSURE tile. A new tile will appear with a PRESSURE display and set-point
arrows: