INSTALLATION AND OPERATING INSTRUCTIONS
Error Code
Warning
Message
Indicator
Light
Priority
Tone
Explanation / Appropriate Actions
E0016
BATTERY
ERROR
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Battery failed/error.
Specifi cation:
The unit has received a signal from the battery pack that
indicated the State of Battery Health is poor. The unit has shut down in
a safe state.
Explanation/Action:
Cycle the ON/STANDBY button then verify the
battery health in the Service Info feature. Replace any unhealthy
batteries. If problem persists, note error code and contact manufacturer.
E0017
BATTERY
ERROR
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Battery failed/error.
Specifi cation:
The unit has received a signal from the battery pack that
indicated the battery has failed. The unit has shut down in a safe state.
Explanation/Action:
Replace the battery with a new Zimmer battery
pack. Cycle the ON/STANDBY button. If problem persists, note error
code and contact manufacturer.
E0024 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Software has detected the potential for instability.
Specifi cations:
Internal system diagnostics feature has detected an
error. The unit has shut down in a safe state.
Explanation/Action:
Cycle the ON/STANDBY Button. If problem
persists, note error code and contact manufacturer.
E0025 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Amplifi er voltage failure detected.
Specifi cation:
The unit has detected a failure in the amplifi er circuit
reference voltage. The unit has shut down in a safe state.
Explanation/Action:
Transducer reference voltage is out of range.
Cycle the ON/STANDBY Button. If problem persists, note error code and
contact manufacturer.
E0026 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Amplifi er voltage error detected in the reservoir.
Specifi cation:
The unit has detected a failure in the amplifi er circuit
reservoir voltage. The unit has shut down in a safe state.
Explanation/Action:
Reservoir Transducer signal voltage is out of range.
Cycle the ON/STANDBY button. If problem persists, note error code and
contact manufacturer.
E0027 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Amplifi er voltage error detected in the Main Cuff.
Specifi cation:
The unit has detected a failure in the amplifi er circuit Main
Cuff voltage. The unit has shut down in a safe state.
Explanation/Action:
Main Cuff Transducer signal voltage is out of range.
Cycle the ON/STANDBY button. If problem persists, note error code and
contact manufacturer.
E0028
XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Amplifi er voltage error detected in the Second Cuff.
Specifi cation:
The unit has detected a failure in the amplifi er circuit
Second Cuff voltage. The unit has shut down in a safe state.
Explanation/Action:
Second Cuff Transducer signal voltage is out of
range. Cycle the ON/STANDBY button. If problem persists, note error
code and contact manufacturer.
E0029 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Amplifi er bulk voltage failure detected.
Specifi cation:
The unit has detected a voltage failure in the bulk voltage
supply.
Explanation/Action:
Bulk voltage is out of range. Cycle the ON/
STANDBY button. If problem persists, note error code and contact
manufacturer.