Page 17
INSTALLATION AND OPERATING INSTRUCTIONS
•
The PRESSURE display should read “250” (the factory default set-point). The tile will automatically close in
approximately 5 seconds.
•
Within the 5 second time frame, touch the negative left-arrow button
-
to decrease or positive right-arrow
button
+
to increase the PRESSURE set-point.
■
The set pressure can be maintained between 50 mmHg and 400 mmHg in increments of 1 mmHg.
■
If an arrow button is continuously held, the increments will change by 5 mmHg.
•
Touch the Okay button
to immediately remove the displayed tile or wait approximately 5 seconds and it will
automatically close.
•
Repeat
PRESSURE set-point system test
for the blue Second Cuff PRESSURE set-point.
WARNING:
•
When medically necessary to use the
A.T.S.
4000TS in the EXTENDED PRESSURE MODE, ensure the
attached tourniquet cuff and hoses are capable of withstanding the extended pressure. Consult the cuff
instructions for cuff pressure limits. Cuffs that have no specifi cations regarding pressure limits should
not be used with high tourniquet pressures.
•
A yellow confi rmation tile appears if the pressure change exceeds the nominal maximum of 400mmHg.
Touch the Okay button
to confi rm need to enter the extended pressure range and to allow continued
increase of pressure to a maximum of 600mmHg, or touch the Cancel button to remain at 400mmHg.
The cuff status bar turns yellow as an ongoing warning. The Pressure tile will also indicate the set-
point pressure in yellow and the word “extended” will appear to show the pressure exceeds that
recommended.
•
Anytime an asterisk (*) appears below the PRESSURE value in the PRESSURE tile, the data displayed
is the set-point. Set pressure will revert back to the default pressure setting when the unit is set to
STANDBY.
2
TIME set-point system test
•
Touch the red Main Cuff TIME tile and a new tile will appear with a TIME display and set-point arrows:
Содержание A.T.S. 4000TS
Страница 1: ...A T S 4000TS Operator Service Manual REF 60 4000 101 00 ENGLISH ...