Page 53
INSTALLATION AND OPERATING INSTRUCTIONS
Functional and Calibration Check
It is recommended that the functional and calibration check described in the
Tests and Checks
section under the
Installation and Operating Instructions
section are performed at least once every three months. A Test and
Analysis form is provided at the back of this manual.
SERVICING
The
A.T.S.
4000TS system has several built-in maintenance/service functions available through a
Service
menu,
including
Calibrate
,
Leak test
,
Unit Info
,
Burn mode
, and
Touch test
.
SERVICE MENU ACCESS
To access the
Service
menu:
•
Touch the
Settings
button
on the Systems Function Menu. The
Settings
menu tile is displayed:
•
Touch and hold the center of the large
Settings
gear symbol fi rst and while continuing to hold the center of the
gear, touch and hold the padlock icon . Continue to hold until the system unlocks the padlock and the
Service
menu is displayed, revealing the
Calibrate
,
Leak test
,
Unit Info
,
Burn mode
, and
Touch test
function buttons.
CAUTION:
•
The Service function should only be accessed by a qualifi ed biomedical engineer or other person
thoroughly familiar with electronic medical devices within the healthcare organization. By default, the
Service button is grayed-out and locked as shown by a padlock icon .
•
If the button combination was not correctly touched and held, the Systems Function Menu will not allow
access and automatically close the menu due to inactivity. While in the Service mode, the system will
keep the function running without automatically closing. The Service mode is for service only and not
for use on or during a procedure.
CALIBRATION
Calibration of the
A.T.S.
4000TS Tourniquet allows the output signal from the pressure transducers to be compared
against a calibrated pressure source. The difference between the known pressure and the pressure measured by
the transducers is recorded at each of fi ve set-points for cuffs (0 mmHg, 50 mmHg, 250 mmHg, 475 mmHg and
600 mmHg) and 5 set-points for the reservoir (0 mmHg, 50 mmHg, 250 mmHg, 475 mmHg, and 700 mmHg). These
calibration points are used to correct the signal from the pressure transducers during normal operation. The calibration
points and a checksum are stored in nonvolatile memory. It is recommended the
A.T.S.
4000TS Tourniquet be
calibrated at least once every 6 months or after any scheduled or unscheduled maintenance.