Operator’s Manual
© Copyright 2004 ZEVEX Incorporated
2. Directions for Use
9
Step 6:
Press and hold
ON/OFF
key for 1.5 seconds
to turn pump on. While pump runs through
a self test, display will light and an audi-
ble alarm will sound as pump displays the nine
digit serial number three digits at a time for
one second each. The next number displayed
for one second, a number beginning with an
"R", is the software revision.
Next, alarm will stop and all segments of dis-
play will be shown for 2 seconds.
Verify all
display segments and symbols are active.
The self test is completed and pump will then
display last programmed rate and will be in
pause mode.
If any air is still in the tubing, continue to
Step 7 to use pump prime feature.
Step 7:
Press and hold
PRIME
key. Alarm will sound
once and pump will begin pumping at the
maximum rate of 600 ml/hr. Display will
read "TO STOP LET GO" (
Figure 2-9).
Once all air
is removed from tubing, release key. Pump will
stop, display will revert to last programmed
rate, and pump will be in pause mode.
For pump operation instructions:
For a Single Feeding Example, go to
Page 10
.
For an Interval Feeding Example, go to
Page 13
.
Figure 2-9
Priming Indication
Figure 2-8 Close Pump Door
Figure 2-7 Seat Cassette
6. Specifications
VOLUME DISPLAY
Current Dose Status
Accumulated
INTERVAL FEED
Interval Feed Indication
Displays Time to Next Feed
PROGRAM MEMORY RETAINED
When Pump is on or off
PEDIATRIC USE
Yes
PRIME
Yes
ALARMS
Chek
Programming error
Load Set
Set not loaded properly
Low Batt
Battery Low
No Flow In
Upstream occlusion
-5 psi (-34 kPa)
Tolerance = ± 3psi (21 kPa)
psi (pounds per square inch)
kPa (kilopascals)
No Flow Out
Downstream occlusion,
12 psi (83 kPa)
Tolerance = ± 3psi (21 kPa)
psi (pounds per square inch)
kPa (kilopascals)
No Food
Bag is empty
ERRA - ERRZ
Self-test error