Operating the Pump
A Single Feeding Example
If you want to deliver 500 ml of enteral feeding solution at a rate of 120 ml/hr
and then stop the pump:
Step 1:
Prepare the disposable set with enteral feeding solution as described in the
Priming and Loading the Disposable Set instructions
(page 7)
.
Step 2:
Hang feeding bag or container so that the bottom of bag is at or above the
level of the pump door.
OR -
If an EnteraLite
®
Infinity
™
carry pack is to be used, load pump and
feeding bag into the proper compartments, securing pump, bag and tubing
with the pack's straps
(Page 30)
.
Step 3:
Turn pump on by pressing the
ON/OFF
key. After running
the self test, display will show last programmed rate.
Step 4:
Press the
+
or
-
key to change the rate to 120 ml/hr.
Hold down either key to change rapidly (
Figure 2-10).
Step 5:
Press the
RATE/DOSE
key to display dose. Press the
+
or
-
key until a dose of 500 ml is displayed (
Figure 2-11).
2. Directions for Use
10
EnteraLite
®
Infinity
™
Operator’s Manual
© Copyright 2004 ZEVEX Incorporated
31
Key Sequence for “A Single Feeding Example”
120 ml/h
500 ml
Figure 2-10 Program
Rate to 120 ml/hr
Figure 2-11 Program
Dose to 500 ml
6. SPECIFICATIONS
6. Specifications
SIZE
Pump Dimensions:
1.95” H x 5.65” W x 4.05” D
Pump Weight:
14.4 oz (411.0 g)
BATTERY
Type:
Lithium Ion
Life:
24 hours @ 125 ml/hr
Charge Time:
approximately 6 hours
Charge Level Indicator Yes
Compact Wall Charger
Yes
DISPLAY
Backlit LCD
OPERATING ORIENTATION
Any
FLOW RATE
Range:
0.1 - 600 ml/hr
Increment:
0.1 ml/hr from 0.1 ml/hr to 9.9 ml/hr
1 ml/hr from 10 ml/hr to 600 ml/hr
Accuracy:
± 5%
DOSE
Range:
0.1 ml to 3000 ml or infinite dose
Increment:
0.1 ml from 0.1 - 9.9 ml
1 ml from 10 - 999 ml
10 ml from 1000 - 3000 ml
25356_pages.qxd 2/2/2005 2:50 PM Page 16