2. Directions for Use
8
EnteraLite
®
Infinity
™
Operator’s Manual
© Copyright 2004 ZEVEX Incorporated
33
NOTE: If set is to be used with a carry pack, all
air must be removed from bag and tubing.
Continue to step 3 for instructions on
removing air
. If set is to be hung above pump,
i.e. on an IV pole,
you may skip to step 4
.
Step 3:
Turn bag upside down and gently squeeze. Tilt bag
as needed to keep air at tubing port (
Figure 2-4).
Step 4:
Gently pinch teal colored tubing
below
“ “
symbol. Hold this position until air is removed
from tubing. Gently squeeze bag at same time
to assist fluid flow. If fluid does not flow, pinch
may be too tight (
Figure 2-5).
NOTE: Inside the teal colored tubing, below the
“ “ symbol is the in-line occluder. By pinching
tubing gently, the tubing moves away from the
in-line occluder allowing fluid to flow (
Figure 2-
6)
. It is important to only pinch the tubing
below the “ “ symbol to avoid damaging the
in-line occluder
.
NOTE: Air may also be removed from tubing
using the pump prime feature. See step 7 for
instructions on using the pump prime feature.
Step 5:
Loop silicone tubing around pump wheel
stretching lightly. Seat cassette into pump
(
Figure 2-7).
Close pump door
(
Figure 2-8).
Figure 2-4 Squeeze Bag
Figure 2-5 Pinch Tubing
Tubing
Port
Figure 2-6 Tubing Segment
Being Pinched
6. Specifications
DISPOSABLE SET FEATURES
Automatic Free-flow Protection
No Drip Chamber
DEHP-free
ACCESSORIES
AC Adapter/Charger
Multi-position Pole Clamp (rotatable 360°)
Multi-purpose EnteraLite
®
Infinity
™
Carry Packs
WARRANTY
2 years
OPERATING ENVIRONMENT
Temperature:
41 °F to 104 °F (5 °C to 40 °C)
Humidity:
30% to 75% noncondensing
STORAGE ENVIRONMENT
Temperature:
-4 °F to 149 °F (-20 °C to 65 °C)
Humidity:
10% to 95% noncondensing
CLASSIFICATION INFORMATION
EnteraLite
®
Infinity
™
Enteral Feeding Pump with internal battery –
Type BF Applied Part
EnteraLite
®
Infinity
™
AC Adapter / Battery Charger – Class II
25356_pages.qxd 2/2/2005 2:50 PM Page 14