28
EnteraLite
®
Infinity
™
Operator’s Manual
© Copyright 2004 ZEVEX Incorporated
2. Directions for Use
Next Feeding:
Turn pump on. Previous settings for rate, dose, and feed interval will be
saved in memory. If no changes are required, verify each setting is correct
and restart pump running.
NOTE: Rate, dose, and feed interval settings can be locked so that changes
cannot be made. See the User Preference Settings section (
Page 19
) for
instructions to change this feature.
Operating the Pump
An Interval Feeding Example
If you want to deliver 100 ml of enteral feeding solution at a rate of 50 ml/hr,
and repeat this feeding every 6 hours:
Step 1:
Prepare the disposable set with enteral feeding solution as described in the
Priming and Loading the Disposable Set instructions
(Page 7)
.
Step 2:
Hang feeding bag or container so that the bottom of bag is at or above the
level of the pump door.
OR -
If an EnteraLite
®
Infinity
™
pack is to be used, load the pump and feed-
ing bag into the proper compartments, securing pump, bag and tubing with
the pack's straps
(see Page 30)
.
Step 3:
Turn pump on by pressing the
ON/OFF
key. After running the self test, dis-
play will show last programmed rate.
13
Key Sequence for “An Interval Feeding Example”
50 ml/h
100 ml
6.00 hr
4. CLEANING
4. Cleaning
WARNING:
To avoid electrical shock, never clean pump or EnteraLite
®
Infinity
™
AC Adapter/Charger with charger plugged into an outlet or pump turned on.
WARNING:
Make sure the EnteraLite
®
Infinity
™
AC Adapter/Charger is com-
pletely dry before plugging into an electrical outlet.
To clean the EnteraLite
®
Infinity
™
Enteral Feeding Pump:
Pump may be cleaned with warm soapy water and
a nonabrasive sponge or soft cloth.
OR -
Apply one of the following solutions for
approximately 10 minutes, then wipe pump clean
with a damp cloth or sponge:
5% bleach and water solution
Multipurpose disinfectant cleaner
You may also rinse the pump by holding it under
a stream of warm water. Then, dry with a clean cloth.
Use a cotton swab to gently clean pathways of cassette receptacle.
Compartments of the EnteraLite
®
Infinity
™
pack can be cleaned with a damp
sponge. The EnteraLite
®
Infinity
™
pack is machine-washable. Use warm water
and gentle cycle, whenever possible. Hang to dry.
EnteraLite
®
Infinity
™
AC Adapter/Charger normally does not require cleaning.
When desired, a dry or slightly damp cloth may be used to clean the outside
surface of AC Adapter/Charger while it is disconnected from the wall outlet.
Figure 4-1 Rinsing Pump
Under Stream of Water
25356_pages.qxd 2/2/2005 2:50 PM Page 19