2. Directions for Use
16
EnteraLite
®
Infinity
™
Operator’s Manual
Feeding Completion:
When the dose has been completely delivered, pump
will stop running and display will read NEXT DOSE then
give the number of hours and minutes until feeding
will be repeated (
Figure 2-23
). Display will show the
run symbol flashing without the arcs rotating around
it. Pump will repeat this cycle until disposable set is
empty or until it is stopped by user.
To stop feeding, press the
ON/OFF
key for 1.5 seconds
to turn pump off or press the
RUN/PAUSE
key to put
pump in pause mode. Adjustments to settings can be
made at this time.
New Feeding Cycle:
Turn pump on. Previous settings for rate, dose, and feed interval will be
saved in memory. If no changes are required, verify each setting is correct
and restart pump running.
NOTE: Rate, dose, and feed interval settings can be locked into place so that
changes cannot be made. See the User Preference Settings section (
Page 19
)
for instructions to change this feature.
© Copyright 2004 ZEVEX Incorporated
25
Figure 2-23 Time Until
Next Feeding
3. Alarms, Messages, and Indications
PUSH RUN TO FEED
(
Figure 3-7
)
Pump will sound a dual tone beep repeatedly and dis-
play will read PUSH RUN TO FEED.
Why:
This alarm has occurred because pump has been
in pause mode for 2 minutes.
To correct the problem:
Press the
RUN/PAUSE
key to
silence alarm and place pump in pause mode again.
Program and use pump or press the
ON/OFF
key to
turn it off.
Messages
DOSE DONE
(
Figure 3-8
)
This message indicates a single feeding dose has been
delivered.
To clear pump:
Either press the
ON/OFF
key to turn
pump off, or press the
RUN/PAUSE
key to put pump in
pause mode before starting another feeding.
NEXT DOSE
(
Figure 3-9
)
If the interval feeding feature is in use, between feed-
ings display will read NEXT DOSE ##.## hr. (##.##,
indicates the time in hours and minutes until the next
dose begins.)
If the currently programmed feeding is not desired,
either press the
ON/OFF
key to turn pump off or press
the
RUN/PAUSE
key to place pump in pause mode
where adjustments can be made to the settings.
Figure 3-7 PUSH RUN
TO FEED Alarm
Figure 3-8
DOSE DONE Message
Figure3-9
NEXT DOSE Message