2. Directions for Use
14
EnteraLite
®
Infinity
™
Operator’s Manual
Step 4:
Press the
+
or
-
key to change the rate to 50 ml/hr.
Hold down either key to change rapidly (
Figure 2-17
).
Step 5:
Press the
RATE/DOSE
key to display dose. Press the
+
or
-
key until a dose of 100 ml is displayed (
Figure 2-18
).
Step 6:
Press the
FEED INT
key to display the feed interval
setting. Press the
+
or
-
key until 6.00 hr is displayed
(
Figure 2-19
).
NOTE: The amount of time to complete a rate and dose
combination can be figured by dividing the dose by
the rate (Dose/Rate = Time). The number of hours to
be programmed as the Feed Interval is the amount of
time to deliver the dose plus the amount of time pump
should pause. Example: 100 ml divided by 50 ml/hr
equals 2 hours for feeding delivery, plus a pause of 4
hours equals a feeding interval of 6 hours (100 ml ÷ 50ml/hr = 2 hours run-
ning + 4 hours pause = 6 hours from feeding start time to start time).
Pump will calculate number of hours it will take to deliver the RATE and
DOSE combination. When the
FEED INT
key is pressed the first available
value will be compatible with that combination. In this example the first
available value will be 2.00 hr.
Step 7:
Connect barbed adapter to patient's enteral feeding tube.
© Copyright 2004 ZEVEX Incorporated
27
Figure 2-17 Program
Rate to 50 ml/hr
Figure 2-18 Program
Dose to 100 ml
Figure 2-19 Program
Feed Interval 6.00 hr
3. Alarms, Messages, and Indications
Battery Doesn't Hold Its Charge
Battery will lose its charge if stored for a long period of time or if stored at
high temperatures (such as in an automobile during summer).
To correct the problem:
Recharge the battery. If the problem continues,
contact your healthcare provider for service.
Battery Doesn’t Charge
If the battery symbol and the E and F of
the fuel gauge
(Figure 3-12)
are flashing
while the pump is plugged in, the battery
is not charging.
To correct the problem:
Contact your
healthcare provider for service.
Charger Installed but No Plug Symbol Visible
If the plug symbol does not display after the AC Adapter/Charger is plugged
into pump and a live power outlet, then charger is not charging battery.
To correct the problem:
Check the following:
Verify that the wall outlet works by plugging in another appliance,
such as a lamp.
Charger is connected properly.
If this does not correct the problem, pump or charger may be damaged.
Contact your healthcare provider for service.
NOTE: There are no user serviceable or replaceable components inside the
EnteraLite
®
Infinity
™
.
Figure 3-12 Battery Symbol and
the E and F of the Fuel Gauge
25356_pages.qxd 2/2/2005 2:50 PM Page 20
To Purchase, Visit
or call